111 In Zevalin Imaging: Purpose
• To verify the expected biodistribution
• R/O rare cases of altered biodistribution
• Not to diagnose the extent of NHL
Data Acquisition for 111 In Zevalin
• Dual or single head gamma camera system
• Medium energy collimator(s)
• 172 and 247 keV with 15% window
• 256 x 1024 matrix
• Speed (dependent on time after injection)
• 10 cm/min (20 minute scan) at 2 - 24 hours
• 7 - 10 cm/min (30 minute scan) at 48 - 72 hours
• 5 - 7 cm/min (40 minute scan) at 90 - 120 hours (optional)
Expected Biodistribution

| Anterior |
Posterior |
Anterior |
Posterior |
Anterior |
Posterior |
4-6 Hrs |
64-70 Hrs |
138-142 Hrs |
Representative Patient: 111 In-Labeled Zevalin® Monoclonal Antibody Imaging

Abdominal SPECT

Abdominal CT

Altered Biodistribution Image

Median 90 Y Zevalin Radiation: Absorbed Doses to Organs and Tumor

|